Efficient 3D FISP-MRF at 0.55 T using long spiral readouts and concomitant field effect mitigation.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Nam G Lee, Krishna S Nayak, Zhibo Zhu

Ngôn ngữ: eng

Ký hiệu phân loại: 621.3815284 Electrical, magnetic, optical, communications, computer engineering; electronics, lighting

Thông tin xuất bản: Netherlands : Magnetic resonance imaging , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 200023

PURPOSE: To demonstrate the feasibility of SNR-efficient 3D fast imaging with steady state precession MR fingerprinting (FISP-MRF) using long spiral readouts with mitigation of concomitant field effects at 0.55 T. METHODS: Fourteen FISP-MRF sequences with different spiral readout lengths (2.9 ms to 22.0 ms) were implemented with the open-source Pulseq framework. Datasets were reconstructed using a low-rank and subspace model-based reconstruction combined with MaxGIRF spatial encoding model. Concomitant field-induced blurring and MRF precision were evaluated using reconstructed images and quantitative maps acquired from an ACR phantom, an ISMRM/NIST system phantom, and 2 healthy volunteers. RESULTS: A simulation study shows that the SNR of time-series images would increase by ~2× in white matter as a spiral readout increased from 2.9 ms to 22.0 ms. Empirically, MRF T CONCLUSION: We demonstrate SNR-efficient 3D FISP-MRF using long spiral readouts in conjunction with concomitant field-induced blurring mitigation. A wide operating regime with improved precision is feasible only after concomitant field mitigation.