Validity and between-days reliability of two different metabolic systems for measuring gas exchange during walking.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Juan M A Alcantara, Fernando Alonso-Gonzalo, Antonio Clavero-Jimeno, Andres Marmol-Perez, Jonatan R Ruiz, Jose Antonio Suarez-Roman

Ngôn ngữ: eng

Ký hiệu phân loại: 371.26013 School administration; administration of student academic activities

Thông tin xuất bản: England : Medical engineering & physics , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 228117

 BACKGROUND: Portable metabolic systems for assessing gas exchange outside the laboratory and clinical settings are increasing their popularity. We aimed to determine the validity and the reliability of the K5 portable system (Cosmed, Rome, Italy) and the Omnical (Maastricht Instruments, Maastricht, The Netherlands) metabolic cart. METHODS: Gas exchange was measured, using the K5 and the Omnical in a randomised and counterbalanced order, in 35 participants (19 women
  age: 26 ± 3 yrs.) using a walking protocol on 2 non-consecutive (48 h apart) days. The protocol consisted of a 5 min walk and 2.5 min rest cycle starting at 3.5 km/h and increasing in 1 km/h to 6.5 km/h. The protocol was repeated twice after a 10-15 min resting. To determine validity and reliability, we conducted two-factor repeated measures analysis of variance (ANOVA), intraclass correlation coefficients (ICC), coefficients of variation (CV) and Bland-Altman analyses. RESULTS: Repeated measures ANOVA showed that mean oxygen consumption, carbon dioxide production and energy expenditure were higher for the K5 compared to the Omnical on both visits (all P <
  0.001). Respiratory exchange ratio (RER) was similar among systems at faster (≥4.5 km/h) walking speeds. ICC, CV and Bland-Altman showed a better reliability of the Omnical (ICCs ranged from 0.71 to 0.86, and CVs from 3.1 % to 7.1 %) compared to the K5 (ICCs ranged from 0.40 to 0.98, and CVs from 2.6 % to 6.3 %). CONCLUSIONS: The K5 portable system provides similar RER values to the Omnical metabolic cart, although the Omnical reliability was better for most outcomes.