Gynecologic cancers are often asymmetric, yet current Ir-192 brachytherapy techniques provide only limited radial modulation of the dose. The shielded solutions investigated here solve this by providing the ability to modulate between highly asymmetric and radially symmetric dose distributions at a given location. To find applicator designs that can modulate between full dose and less than 50% dose, at the dimensions of the urethra, a 2D calculation algorithm was developed to narrow down the search space. Two shielding design types were then further investigated using Monte Carlo and Boltzmann-solver dose calculation algorithms. 3D printing techniques using ISO 10993 certified biocompatible plastics and 3D printable tungsten-loaded plastics were tested. It was also found that shadowing effects set by the shape of the shielding cannot be easily modulated out, hence careful design is required. The shielded applicator designs investigated here, allow for reduction of the dose by over 50% at 5 mm from the applicator surface in desired regions, while also allowing radially symmetric dose with isodose line deviations less than 0.5 mm from circular. The shielding designs were also chosen with treatment delivery time in mind. Treatment times for these shielded designs were found to be less than 1.4 times longer than a 6-channel unshielded cylinder for the equivalent fully symmetric dose distribution. The 2D calculation methods developed here provide a simple way to rapidly evaluate shielding designs, while the 3D printing techniques also allow for devices with novel shapes to be rapidly prototyped. Both TOPAS Monte Carlo and Acuros BV calculations show that significant dose shaping and organ at risk sparing can be achieved without significantly compromising the plan in regions that require the full dose.