In this study, we presented the method for calculation the absolute dose in the Monte Carlo simulation following the prescription of Popescu et al for the 6 MV photon energy. The BEAMnrc was used to simulate 6 MV photon beams from a Siemens Primus M5497 linear accelerator at DongNai general hospital. The DOSXYZnrc was then used to calculate the dose distribution in a homogeneous phantom (in form of CT images). The absolute dose obtained from the MC and TPS were compared with measured ones using an ion chamber (Farmer Type Chamber FC65-P, IBA). The average doses discrepancy between the simulated and measured dose was 0.53±0.37% and between the simulated and TPS was 1.00±0.51%. Results showed good agreement between simulated, measured and calculated dosed on a homogeneous phantom.Trong nghiên cứu này, chúng tôi trình bày phương pháp tính toán liều tuyệt đối trong mô phỏng Monte Carlo (MC) áp dụng công trình nghiên cứu của Popescu và cộng sự cho mức năng lượng photon 6 MV. Chương trình BEAMnrc được sử dụng để mô phỏng chùm photon 6 MV phát ra từ máy gia tốc tuyến tính Siemens Primus M5497 tại Bệnh viện Đa khoa Đồng Nai. Sau đó chương trình DOSXYZnrc được sử dụng để tính toán phân bố liều trên ảnh cắt lớp (computed tomography - CT) của phantom đồng nhất. Các giá trị liều từ mô phỏng MC và phần mềm lập kế hoạch (TPS) được so sánh với các kết quả đo đạc bằng thực nghiệm sử dụng buồng ion hóa FC65-P. Sai khác trung bình giữa mô phỏng với các giá trị liều đo đạc và tính trên TPS tương ứng là 0,33 0,15% và 1,00 0,51%. Các kết quả cho thấy có sự phù hợp tốt giữa liều mô phỏng, đo đạc và tính toán trên phantom đồng nhất.