Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Brett G Erickson, Joseph P Kowalski, Xinyi Li, Qiuwen Wu, Sua Yoo

Ngôn ngữ: eng

Ký hiệu phân loại: 658.32259 Personnel management (Human resource management)

Thông tin xuất bản: United States : Journal of applied clinical medical physics , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 549100

 This article investigates the performance of a commercial BeO optically stimulated luminescent (OSL) dosimetry system (myOSLchip, RadPro GmbH International, Remscheid, Germany) through the application of the commissioning framework for luminescent dosimeters as described in the American Association of Physicists in Medicine Task Group 191 (AAPM TG191) report. Initial clinical experiences and dosimetric results are also presented. The following properties of the system were characterized: linearity correction factors ranged from -0.5% to +3% for dose levels spanning 0.1 to 20 Gy. Beam quality correction factors (relative to 6 MV) ranged from -4.5% (2.5FFF) to +4.5% (15MV) for photon beams and +1.9% (6 MeV) to +4.3% (20 MeV) for electron beams. An average (µ) signal loss per reading of -2.13% ± 0.20% was measured, however greater signal loss was observed in the first reading (µ = -2.6% ± 0.46%). An initial decline in individual element sensitivity relative to baseline was observed from 0-15 Gy cumulative dose (µ = -1.98% ± 0.55%), with negligible further deterioration from 15-32 Gy (µ = -2.38% ± 0.85%). Post-irradiation, there was a transient OSL signal which faded with a half-life of 1.8 min
  this signal enhancement was +5% at 5 min post-irradiation and +1% at 15 min relative to 24 h. Dosimeter response was not dependent on average dose rate in the range of 100-2500 MU/min. With respect to clinical testing, equal or superior performance compared with aluminum oxide OSLs (nanoDots) is shown for a range of clinical techniques and modalities including TSET, TBI, en-face electrons, and pacemaker/out-of-field measurements. The feasibility of myOSLchip to serve as a primary clinical in vivo dosimetry system and direct replacement for Landauer's microStar system is demonstrated.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH