PURPOSE: To evaluate the diagnostic validity of the HNT-1P non-contact tonometer (Huvitz) as a tool for accurately measuring intraocular pressure (IOP) in patients with healthy eyes, compared to the Goldmann applanation tonometer (GAT), which is the Gold Standard method for measurement of IOP. METHODS: Observational, descriptive, transversal study using 148 eyes of 74 healthy patients without a diagnosis of glaucoma or other ophthalmological diseases. Three measurements of IOP were taken in each eye, using three tonometers: HNT-1P, ICR100, and GAT. The median IOP (quartiles) and mean IOP (SD) its statistical significance were calculated, and comparisons were made between the mean and median IOP values found in three groups: GAT-HNT, GAT-ICR, and HNT-ICR. The difference in mean and median IOP was analyzed in each of the three study groups, and its statistical significance and concordance correlation coefficient (CCC) were calculated. RESULTS: The median IOP with HNT-1P was statistically significantly lower than the median IOP with GAT, (1.1 mmHg, CONCLUSIONS: HNT-1P (Huvitz) provides statistically significantly lower IOP values than those obtained with GAT. HNT-1P could be used for screening of ocular hypertension in postoperative patients. The IOP measurement obtained with HNT-1P should be confirmed with GAT. HNT-1P yields lower IOP values than those obtained with ICR.