Establishing Decisional Cutoff Values of Neurofilament Light Chains in Cerebrospinal Fluid Measured by Fully Automated Chemiluminescent Enzyme Immunoassay.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Luisa Agnello, Anna Maria Ciaccio, Marcello Ciaccio, Fabio Del Ben, Caterina Maria Gambino, Concetta Scazzone

Ngôn ngữ: eng

Ký hiệu phân loại: 133.594 Types or schools of astrology originating in or associated with a

Thông tin xuất bản: United States : Journal of clinical laboratory analysis , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 641618

 INTRODUCTION: Neurofilament light chain (NfL) is one of the most important biomarkers in the field of clinical neurochemistry. Several analytical methods have been developed in the last decade. Recently, Fujirebio introduced a ready-to-use assay kit for measuring NfL levels in the cerebrospinal fluid (CSF) on the fully automated LUMIPULSE G System. In this study, we established the decisional cutoffs for CSF NfL. MATERIALS AND METHODS: We performed a retrospective observational study including patients with cognitive decline. CSF NfL levels were measured by two analytical methods: the NF-light ELISA kit (UmanDiagnostics) and the Lumipulse G1200 fully automated system (Fujirebio). We calculated the cutoffs for the Lumipulse, starting from the consolidated cutoffs of the ELISA method for each age and using the equation obtained by the regression analysis. RESULTS: The study population consisted of 100 patients with cognitive decline. The median levels of CSF NfL measured by Lumipulse and ELISA were 776.5 ± 772.6 pg/mL and 473.5 ± 443.5 pg/mL, respectively, significantly different (p <
  0.001). The Spearman's rank correlation coefficient was 0.962, indicating a robust positive correlation between the two measurement methods. The equation derived from the Passing-Bablok regression analysis was CSF CLEIA = -61.16 + 1.83 × CSF ELISA. Based on this equation, we defined the decisional cutoff values. CONCLUSIONS: Decisional cutoffs are fundamental tools for guiding clinicians to use biomarkers' results and interpretation appropriately. This is the first study establishing the decisional cutoff value of NfL measured by Lumipulse, a fully automated platform widely used in clinical laboratories.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH