System Accuracy and Interference Evaluation of a New Glucose Dehydrogenase-Based Blood Glucose Meter for Patient Self-Testing.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Mina Hanna, Daiva Kalasauske, Andreas Pfützner, Gerhard Raab, Daniela Sachsenheimer, Nicole Thomé, Silvia Weissenbacher

Ngôn ngữ: eng

Ký hiệu phân loại: 535.47 Interference

Thông tin xuất bản: United States : Journal of diabetes science and technology , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 677050

 New European medical device regulations require the performance of postmarketing surveillance evaluations for blood glucose meters (BGMs). We conducted an ISO15197:2015-conform system performance evaluation with the approved glucose dehydrogenase (GDH)-based Wellion NEWTON BGM. One hundred subjects were enrolled into the study (44 female, 56 male, 43 healthy subjects, 23 type 1 diabetes, 34 type 2 diabetes, age: 53.7 ± 15.8 years). In addition, manipulated heparinized whole blood was used for a laboratory interference test with ten selected substances (interference definition: substance-induced bias >
  10%). The mean absolute relative difference (MARD) was 4.7%, and 100% of the values were in zones A (99.7%) and B (0.3%), respectively, of the consensus error grid. Interference was observed with xylose only, which is a known interfering substance for GDH-based BGMs.