Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Melanie Koh, Afrina Adlyna Mohamad Matrol, Hang Cheng Ong, Letchumy Praba Ramanaidu, Shier Nee Saw, Wei Ping Eddy Tan

Ngôn ngữ: eng

Ký hiệu phân loại:

Thông tin xuất bản: United States : Journal of visualized experiments : JoVE , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 684669

 This study aims to validate the accuracy of low-cost fitness smartwatches by comparing their data with gold-standard measurements for cardiovascular and physical activity parameters. The study enrolled 50 subjects, 26 undergoing validation testing for heart rate, blood oxygen saturation (SpO2), and sleep data against polysomnography (PSG). Additionally, 24 subjects participated in the 3-Minute Walk Test (3MWT) and Stairs Climbing (SC), with step counts validated against manual video calculations. Results showed no significant difference between the device's measurements and gold standard values for shallow sleep, deep sleep, REM time, mean heart rate, minimum heart rate, and SpO2. However, the device significantly underestimated manually counted steps (p = 0.009 (3MWT)
  p = 0.012 (SC)), total sleep duration (p = 0.004), and wake time (p = 8.94 × 10