Direct Visualization of External Outflow Graft Obstruction in Durable Left Ventricular Assist Device Using Transthoracic Echocardiography.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Hiroaki Hiraiwa, Shingo Kazama, Shotaro Komeyama, Toru Kondo, Ryota Morimoto, Toyoaki Murohara, Masato Mutsuga, Shin Nagai, Takahiro Okumura, Tomo Yoshizumi

Ngôn ngữ: eng

Ký hiệu phân loại: 628.16723 Water supply

Thông tin xuất bản: United States : ASAIO journal (American Society for Artificial Internal Organs : 1992) , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 684852

External outflow graft obstruction (OGO) in durable left ventricular assist devices (LVADs) is a rare but critical complication. Detecting external OGO can be challenging because of its nonspecific symptoms and a wide variety of differential diagnoses. We report the case of a 45 year old man with LVAD admitted with complaints of recurrent syncope and frequent low-flow alarms. Laboratory tests and chest radiography revealed no specific findings. Transthoracic echocardiography via the left subcostal approach revealed an outflow graft compressed by a low-echo area within a bend relief. This finding led to a diagnosis of external OGO, which was consistent with what was observed on contrast-enhanced computed tomography. The patient underwent surgery and the gelatinous substance compressing the outflow graft was successfully removed. This case highlights the utility of transthoracic echocardiography as a noninvasive and effective method for detecting external OGOs, emphasizing its role as a valuable tool for the initial screening of patients with suspected OGO.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH