Using the Seraph® 100 Microbind® Affinity blood filter under slow flow conditions through 18 G and 16 G central lines.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Dan-Nicolae Borchina, Jan T Kielstein, Torsten M Meyer, Julius J Schmidt, Malin-Theres Seffer

Ngôn ngữ: eng

Ký hiệu phân loại: 629.1325214 Aerospace engineering

Thông tin xuất bản: United States : The journal of vascular access , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 687964

 INTRODUCTION: The Seraph® 100 Microbind® Affinity blood filter (Seraph® 100) has been in use since 2019 for the treatment of fulminant or difficult to treat blood stream infections as an adjunct to pharmacotherapy. In 2020 the device received emergency use authorization by the US Food and Drug Administration for the treatment of critically ill COVID-19 patients with confirmed or imminent respiratory failure. Results of an international registry showed that the Seraph® 100 was operated under blood flow rates of 100-350 mL/min. As those conditions require a large bore central line, a dialysis catheter is currently considered indispensable to operate the Seraph® 100. The use of smaller catheter lumina has neither been evaluated in vitro nor in vivo. METHODS: In vitro pressure data before and after the Seraph® 100 at various blood pump rates (prepump line 16 G, postpump line 18 G) with saline and human plasma were obtained. Further, anecdotal flow and pressure data of two patients treated with the Seraph® 100 for a COVID-19 infection are reported. RESULTS: At a pump speed of 50 mL/min pre-Seraph® pressure using saline was -70 [-70 to -60] mm Hg. In comparison, using plasma pre-Seraph® pressure was lower at -120 [-120 to -105] mm Hg
  CONCLUSION: Seraph® 100 blood flow rate of 50 mL/min may be achieved using low flow vascular access, allowing to treat a blood volume 72 L in 24 h.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH