Quantitative MRI for Monitoring Metabolic Dysfunction-Associated Steatotic Liver Disease: A Test-Retest Repeatability Study.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Naim Alkhouri, Anneli Andersson, Ami Banerjee, Cayden Beyer, Kathleen Corey, Andrea Dennis, Mukesh Harisinghani, Prashant Pandya, Michele Pansini, Elizabeth Shumbayawonda

Ngôn ngữ: eng

Ký hiệu phân loại: 631.847 Biological methods of soil nitrification

Thông tin xuất bản: United States : Journal of magnetic resonance imaging : JMRI , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 688331

 BACKGROUND: Quantitative magnetic resonance imaging metrics iron-corrected T1 (cT1) and liver fat from proton density fat-fraction (PDFF) are both commonly used as noninvasive biomarkers for metabolic dysfunction-associated steatohepatitis (MASH)
  however, their repeatability in this population has rarely been characterized. PURPOSE: To quantify the variability of cT1 and liver fat fraction from PDFF in patients with biopsy-confirmed metabolic dysfunction-associated steatotic liver disease (MASLD) and MASH. STUDY TYPE: Prospective, single center. POPULATION: Twenty-one participants (female = 11, mean age 53 ± 24 years) with biopsy-confirmed MASLD, including 6 with MASH and fibrosis ≥2. FIELD STRENGTH/SEQUENCE: 3 T
  T1 and T2* mapping for the generation of cT1 (shMOLLI: CardioMaps and 2D MDE, T1map-FIESTA and LMS MOST: StarMap, 2D Multi-Echo FSPGR) and magnitude-only PDFF sequence for liver fat quantification (LMS IDEAL: StarMap, 2D Multi-Echo FSPGR). ASSESSMENT: T1 mapping and PDFF scans were performed twice on the same day for all participants (N = 21), with an additional scan 2-4 weeks later for MASH patients with fibrosis ≥2 (N = 6). Whole liver segmentation masks were generated semi-automatically and average pixel counts within these masks were used for the calculation of cT1 and liver fat fraction. STATISTICAL TESTS: Bland-Altman analysis for repeatability coefficient (RC) and 95% limits of agreement (LOA) and intraclass correlation coefficient (ICC). RESULTS: Same-day RC was 32.1 msec (95% LOA: -36.6 to 24.2 msec) for cT1 and 0.6% (95% LOA: -0.5% to 0.7%) for liver fat fraction
  the ICCs were 0.98 (0.96-0.99) and 1.0, respectively. Short-term RC was 65.2 msec (95% LOA: -63.8 to 76.5 msec) for cT1 and 2.6% (95% LOA: -2.8% to 3.1%) for liver fat fraction. DATA CONCLUSION: In participants with MASLD and MASH, cT1 and liver fat fraction measurements show excellent test-retest repeatability, supporting their use in monitoring MASLD and MASH. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY: Stage 2.