Performance Comparison of Liquid Chromatography and Paper Spray Ionization with Mass Spectrometry for Measuring Kinase Inhibitors in Human Plasma.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Paul J Jannetto, Lisa A Kottschade, Richard G Lahr, Makenzie Meyer, Leah Nelson, Yifei K Yang

Ngôn ngữ: eng

Ký hiệu phân loại: 543.84 Liquid chromatography

Thông tin xuất bản: United States : ACS pharmacology & translational science , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 691420

 Kinase inhibitors are small-molecule drugs designed to target oncogenic mutations in cancer treatment. Although less toxic than conventional chemotherapy drugs, they can cause severe adverse effects in some patients, resulting in dose reduction and cessation. To evaluate if therapeutic drug monitoring of kinase inhibitors and their metabolites can improve toxicity assessment in patients, we developed and evaluated the analytical performance of two parallel methods utilizing liquid chromatography (LC) and paper spray (PS) ionization coupled with a triple quadrupole mass spectrometer (MS) for the measurement of dabrafenib, its major metabolite OH-dabrafenib, and trametinib in patient plasma samples. The PS-MS method yielded a faster sample analysis time (2 min) compared to the LC separation (9 min). The two methods shared the same analytical measurement range (AMR) for dabrafenib and OH-dabrafenib (10-3500 and 10-1250 ng/mL), but the AMR differed for trametinib (LC-MS: 0.5-50 ng/mL
  PS-MS: 5.0-50 ng/mL). The imprecision across their respective AMR was 1.3-6.5% (dabrafenib), 3.0-9.7% (OH-dabrafenib), and 1.3-5.1% (trametinib) for the LC-MS method and 3.8-6.7% (dabrafenib), 4.0-8.9% (OH-dabrafenib), and 3.2-9.9% (trametinib) for the PS-MS method. Using authentic patient samples, the quantification results were comparable between the two methods: dabrafenib (correlation coefficient
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH