Comparing HemoCue® and Quantitative Buffy Coat® and Coulter Counter-measured haemoglobin concentrations in African children with acute uncomplicated malaria: a Bland-Altman analysis.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Seun Ajayi, Dhol S Ayuen, Natenapa Chimjinda, Nicholas P J Day, Caterina Fanello, Kathryn Maitland, Rita Muhindo, Mavuto Mukaka, Peter Olupot-Olupot, Marie A Onyamboko, Chiraporn Taya, Walter R Taylor, Sophie Uyoga, Thomas N Williams

Ngôn ngữ: eng

Ký hiệu phân loại:

Thông tin xuất bản: England : Malaria journal , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 696775

BACKGROUND: Anaemia is a deleterious consequence of malaria, and its accurate diagnosis is crucial for effective management. However, laboratory methods for measuring haemoglobin (Hb) concentration, like the Coulter Counter and the Quantitative Buffy Coat® (QBC®), are costly and not widely accessible in resource-limited settings. The point-of-care HemoCue® test is a cheaper alternative and suitable in rural areas. The study aimed to determine the level of agreement between Coulter Counter/QBC® vs. HemoCue®-measured Hb concentrations by Bland-Altman analysis. METHODS: As part of a randomized, placebo-controlled trial of single low-dose primaquine in Ugandan and Congolese children with acute uncomplicated Plasmodium falciparum malaria, Hb concentrations were measured on days 0, 3, 7, and 28 using Coulter Counter (Uganda, n = 1880 paired values), QBC® (DR Congo, n = 1984 paired values) and HemoCue® Hb-301™. The predefined clinically acceptable limits were set at ± 0.5 g/dL. RESULTS: The Bland-Altman analysis showed that the HemoCue® minus Coulter Counter mean Hb difference was - 0.15 g/dL with lower and upper limits of agreement of - 3.68 g/dL and 3.39 g/dL, respectively. Corresponding HemoCue® minus QBC® values were - 0.23 g/dL, - 1.66 g/dL and 1.22 g/dL. Linear regression of Hb concentration differences vs. mean Hb concentrations showed negative correlations: r = - 0.43 and r = - 0.34 for HemoCue® vs. Coulter Counter and HemoCue® vs. QBC®, respectively. CONCLUSIONS: Compared to Coulter and QBC®, mean HemoCue® measured Hb concentrations were lower and, compared to the Coulter or QBC® methods, had an overall tendency to measure lower Hb concentrations with increasing Hb concentrations. Upper and lower limits of agreement were wider than the predefined clinically acceptable limits of ± 0.5 g/dL. HemoCue® should be used with caution in settings where decisions about blood transfusions are made.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH