Magnetic field quality conversion factors experimentally measured in clinical MR-linac beams for seven MR-compatible ionization chamber models.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Jennie Crosby, Wesley Culberson, Carri Glide-Hurst, Nathan Orlando, Arman Sarfehnia

Ngôn ngữ: eng

Ký hiệu phân loại: 636.681 Swans

Thông tin xuất bản: United States : Journal of applied clinical medical physics , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 700308

PURPOSE: The purpose of this work was to experimentally quantify MR-compatible ionization chamber response for 1.5T Elekta Unity and 0.35T ViewRay MRIdian MR-linac systems through the determination of the magnetic field quality conversion factor, k METHODS: Seven MR-compatible ionization chamber models from Standard Imaging and PTW were evaluated. Both the quality conversion factor k RESULTS: Beam quality conversion (k CONCLUSIONS: This work provides critical experimental validation of MR-compatible ionization chamber performance, with a direct comparison of measured k