Establishing reference ranges for Century Clot® coagulation and platelet function analyzer in third trimester of pregnancy women: An observational study.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Mingyue Ge, Shaoqiang Huang, Tingting Sun

Ngôn ngữ: eng

Ký hiệu phân loại: 636.0885 Animal husbandry

Thông tin xuất bản: United States : Medicine , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 723601

 This study aimed to establish reference ranges of the Century Clot® coagulation and platelet function analyzer (CPA) for pregnant women at ≥37 weeks of gestation. A total of 300 patients were included in this study. Use Century Clot coagulation and platelet function analyzer to analyze the venous blood of the subjects. Numerical values were recorded and differences in coagulation function indicators between normal full-term pregnant women and healthy non-pregnant women were compared, including activated clotting time (ACT), fibrin coagulation rate (CR), and platelet function (PF). Significant differences were observed in the results of relevant indicators measured using the CPA between women with late pregnancy and healthy non-pregnant women (P <
  .0001). The reference range for parameters of CPA in late pregnancy (pregnancy >
  37 weeks) is determined by percentile method as follows: ACT 75.10 to 112.00 seconds, CR 23.20 to 46.98 mm/minute, PF 3.10 to 5.10 units. Compared with non-pregnant women, ACT in late-pregnancy women detected by the CPA decreased physiologically, while fibrin CR and PF increased. The establishment of a specific reference interval for ACT, CR, and PF in pregnant women is valuable for clinical practice.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH