Beam hardening of K-edge contrast agents: a phantom study comparing clinical energy-integrating detector and photon-counting detector CT systems.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Arnaud Richard Choux, Carlo De Cecco, Tilman Emrich, Thomas Wesley Holmes, Amir Pourmorteza, U Joseph Schoepf, Marly van Assen, Akos Varga-Szemes

Ngôn ngữ: eng

Ký hiệu phân loại:

Thông tin xuất bản: England : European radiology experimental , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 724025

 BACKGROUND: Beam hardening (BH) artifacts negatively influence computed tomography (CT) measurements, especially when due to dense materials or materials with high effective atomic numbers. Photon-counting detectors (PCD) are more susceptible to BH due to equal weighting of photons regardless of their energies. The problem is further confounded by the use of contrast agents (CAs) with K-edge in the diagnostic CT energy range. We quantified the BH effect of different materials comparing energy-integrating detector (EID)-CT and PCD-CT. METHODS: Pairs of test tubes were filled with dense CA (iodine-, gadolinium-, and bismuth-based) and placed inside a water phantom. The phantoms were scanned on EID- and PCD-CT systems, at all available tube voltages for the PCD scanner. Images were reconstructed with standard water BH correction but without any iodine/bone BH corrections. Virtual monoenergetic images (VMI) were calculated from PCD-CT data. RESULTS: PCD-CT had higher CT numbers in all x-ray spectra for all CAs (p <
  0.001) and produced larger cupping artifacts in all test cases (p <
  0.001). Bismuth-based CA artifacts were 3- to 5-fold smaller than those of iodine- or gadolinium-based CA. PCD-CT-based VMI completely removed iodine BH artifacts. Iodine BH artifacts decreased with increasing tube voltage. However, gadolinium-based BH artifacts had a different trend increasing at 120 kVp. CONCLUSION: EID had fewer BH artifacts compared to PCD at x-ray tube voltages of 120 kVp and higher. The inherent spectral information of PCDs can be used to eliminate BH artifacts. Special care is needed to correct BH artifacts for gadolinium- and bismuth-based CAs. RELEVANCE STATEMENT: With the increasing availability of clinical photon-counting CT systems offering the possibility of dual contrast imaging capabilities, addressing and comprehending the BH artifacts attributed to old and novel CT CAs grows in research and ultimately clinical relevance. KEY POINTS: EID-CT provides fewer BH artifacts compared to PCD-CT at x-ray tube voltages of 120 kVp and higher. K-edge CAs, such as those based on gadolinium, further confound BH artifacts. The inherent spectral information of photon counting detector CT can be used to effectively eliminate BH artifacts.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH