OBJECTIVE: The dynamic range (DR) available to the patient is a central parameter to determine speech intelligibility in quiet. DESIGN: In this retrospective study, the DR for the Vibrant Soundbridge implanted in individual patients was calculated using STUDY SAMPLE: A data set of 66 cases (4 bilateral and 2 revised cases) from 60 implanted patients between 14.3-81.8 years were analysed. RESULTS: The relationship between DR and WRS was described by a sigmoidal growth function with R CONCLUSION: A minimum DR of 20 dB can yield sufficient speech intelligibility in quiet in implanted patients, however, an optimum DR is suggested to be 40 dB.