Digital detector PET/CT increases Centiloid measures of amyloid in Alzheimer's disease: A head-to-head comparison of cameras.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Pierrick Bourgeat, Svetlana Bozinovski, Timothy Cox, Vincent Doré, Azadeh Feizpour, Jurgen Fripp, Ashley Gillman, Kun Huang, Natasha Krishnadas, Shenpeng Li, Graeme O'Keefe, Christopher C Rowe, Rosita Shishegar, Victor L Villemagne, Rob Williams

Ngôn ngữ: eng

Ký hiệu phân loại:

Thông tin xuất bản: United States : Journal of Alzheimer's disease : JAD , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 737375

 BACKGROUND: The introduction of therapeutics for Alzheimer's disease has led to increased interest in precisely quantifying amyloid-β (Aβ) burden for diagnosis, treatment monitoring, and further clinical research. Recent positron emission tomography (PET) hardware innovations including digital detectors have led to superior resolution and sensitivity, improving quantitative accuracy. However, the effect of PET scanner on Centiloid remains relatively unexplored and is assumed to be minimized by harmonizing PET resolutions. OBJECTIVE: To quantify the differences in Centiloid between scanners in a paired cohort. METHODS: 36 participants from the Australian Imaging, Biomarker and Lifestyle study (AIBL) cohort were scanned within a year on two scanners. Each participant underwent RESULTS: Centiloid from the Vision camera was found to be significantly higher compared to the Gemini and mCT
  the difference was greater at high-Centiloid levels. Post-reconstruction resolution harmonization only accounted for and corrected ∼20% of the Centiloid (CL) difference between scanners. We further demonstrated that residual differences have effects that vary spatially between different subregions of the Centiloid mask. CONCLUSIONS: We have demonstrated that the type of PET scanner that a participant is scanned on affects Centiloid quantification, even when scanner resolution is harmonized. We conclude by highlighting the need for further investigation into harmonization techniques that consider scanner differences.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH