A Case of Successful Extraction of a Micra Leadless Pacemaker Implanted Two Years Prior Using the Axis-Guided Dual Snare Technique.

 0 Người đánh giá. Xếp hạng trung bình 0

Tác giả: Tetsuhisa Hattori, Ken Kakita, Misun Pak, Takashi Yamasaki

Ngôn ngữ: eng

Ký hiệu phân loại: 616.833 *Parkinson disease

Thông tin xuất bản: United States : Case reports in cardiology , 2025

Mô tả vật lý:

Bộ sưu tập: NCBI

ID: 748464

A 78-year-old man underwent Micra AV implantation due to complete atrioventricular block. He developed diffuse left ventricular systolic dysfunction and dyssynchrony, 2 years later, suggesting pacing-induced cardiomyopathy. Given the critical need for an upgrade to cardiac resynchronization therapy (CRT), an extraction of the Micra AV was scheduled. Initially, the Micra extraction using a solitary snare catheter was attempted, but capturing the retrieval feature on the proximal side of the device was unsuccessful due to excessive movement caused by the heartbeats. Consequently, an axis-guided dual snare technique employing two snare catheters was implemented. A triple-loop snare catheter was inserted into a steerable sheath, which was threaded through the loop of a single-loop snare catheter placed outside the sheath. The triple-loop snare successfully grasped the Micra body and stabilized its movement. Subsequently, the single-loop snare was advanced along the shaft of the triple-loop snare catheter towards the retrieval feature. The single-loop snare finally captured the retrieval feature, enabling the smooth retraction of the Micra into the sheath. Following the successful extraction of the Micra, a CRT device was implanted without complications.
Tạo bộ sưu tập với mã QR

THƯ VIỆN - TRƯỜNG ĐẠI HỌC CÔNG NGHỆ TP.HCM

ĐT: (028) 36225755 | Email: tt.thuvien@hutech.edu.vn

Copyright @2024 THƯ VIỆN HUTECH